The U.S. Food and Drug Administration is notifying health care professionals
and patients that Tyco Healthcare Group LP (Covidien)
is recalling one lot of ReliOn sterile, single-use, disposable, hypodermic
syringes with permanently affixed hypodermic needles due to possible mislabeling.
The use of these syringes may lead to patients receiving an overdose of as
much as 2.5 times the intended dose, which may lead to hypoglycemia, serious
health consequences, and even death.
The recall applies to the following lot number and product information:
-- Lot Number 813900
-- ReliOn 1cc, 31-gauge, 100 units for use with U-100 insulin
Only ReliOn syringes from this lot number and labeled as 100 units for use
with U-100 insulin are the subject of the recall.
These syringes are distributed by Can-Am Care Corp and sold only by Wal-Mart
at Wal-Mart stores and Sam's Clubs under the ReliOn name. Wal-Mart requests
that all users of ReliOn 31-gauge, 1cc syringes return those labeled as 100
units for use with U-100 insulin from Lot Number 813900 to their local Wal-Mart
store or Sam's Club pharmacy. Customers will be provided with replacement
product.
The FDA urges patients and health care professionals to check their syringe
packaging carefully for syringes labeled as 100 units for use with U-100 insulin
from Lot Number 813900.
Consumers and health care professionals who suspect they have the recalled
product may also contact Covidien at 866-780-5436 or www.relion.com/recall for
more information.
ReliOn Insulin Syringes consist of a syringe barrel, a plunger rod, and a
hypodermic needle attached to the tip of the syringe.
During the packaging process for this lot, some syringes labeled for use with
U-40 insulin were mixed with syringes labeled for use with U-100 insulin, then
all packaged individually and in boxes as 100 units for use with U-100 insulin.
The manufacturer has distributed 4,710 boxes in the recalled lot, which equals
471,000 individual syringes. Wal-Mart sold the syringes at Wal-Mart stores
and Sam's Clubs from Aug. 1, 2008, until Oct. 8, 2008.
Tyco Healthcare Group LP (Covidien) voluntarily recalled this lot of syringes
on Oct. 9, 2008, asking that any units of the affected product be removed from
inventory and placed in quarantine. Wal-Mart posted the recall announcement
in Wal-Mart stores and Sam's Clubs, as well as on its Web site, and sent
letters to more than 16,500 customers notifying them of the recall.
The manufacturer has received one adverse report related to a syringe from
this product lot.
Health care professionals and consumers may report serious adverse events (side
effects) or product quality problems with the use of this product to the FDA's
MedWatch Adverse Event Reporting program either online, by regular mail, fax
or phone.
--Online: www.fda.gov/MedWatch/report.htm
--Regular Mail: use postage-paid FDA form 3500 available at: www.fda.gov/MedWatch/getforms.htm and
mail to MedWatch, 5600 Fishers Lane, Rockville, MD 20852-9787
--Fax: (800) FDA-0178
--Phone: (800) FDA-1088
-------------------------------------------------------------------------------------------------------------------------
|
FOR IMMEDIATE RELEASE
November 5, 2008
|
Media
Inquiries:
Siobhan DeLancey, 301-827-6242
Consumer Inquiries:
888-INFO-FDA
|