# Carefusion Provides Update on Voluntary Global Recall of Avea® Ventilator
- URL: https://www.foodconsumer.org/voluntary_global_recall_of_avea_ventilator_0528150852/
- Published: 2015-05-28T09:51:00.000Z
- Updated: 2026-09-15T14:33:56.000Z
- Description: CareFusion says FDA has designated its global AVEA ventilator recall Class I. A 5 psi pressure transducer can fail over time, triggering false alarms, opening the safety valve and stopping ventilation. Units made or serviced between July 2011 and March 2015 are affected; no injuries reported.
- Author: Food Consumer
- Tags: Drug, Health News, #vivvo-import

# Carefusion Provides Update on Voluntary Global Recall of Avea® Ventilator

**Contact:**  
Consumer:  
888-562-6018  
  
Technical Support:  
888-231-2466  
  
Media:  
Troy Kirkpatrick   
(858) 617-2361  
Email: [troy.kirkpatrick@carefusion.com](mailto:troy.kirkpatrick@carefusion.com)

**FOR IMMEDIATE RELEASE** — **May 27, 2015** — **San Diego, CA** — CareFusion, a BD company (NYSE:BDX), provided an update on a global voluntary recall that was initiated on April 21, 2015 to address an issue with certain units of AVEA® ventilators.

The recall has been designated as a Class I recall by the U.S. Food and Drug Administration (FDA), which means the Agency believes that there is a reasonable probability that use of the recalled product will cause serious adverse health consequences or death. To date, no report of patient injury has been received related to this issue.

The AVEA ventilator is only used in hospitals and other health care facilities and is intended for continuous breathing support for neonatal through adult patients. The recall is in response to a potential malfunction of an AVEA ventilator specific 5 psi pressure transducer. The affected AVEA ventilators may develop a failure mode over a period of time, where, by design, the ventilator activates false Extended High Ppeak or Circuit Occlusion audio and visual alarms, opens the safety valve and stops ventilating. If this occurs, alternate ventilation support will be required to reduce the potential of hypoxemia or hypercapnia.

The global recall involves AVEA ventilators manufactured, serviced and distributed from July 1, 2011 to March 15, 2015\. CareFusion learned of this issue through customer reports identifying Extended High Ppeak or Circuit Occlusion alarms. A list of affected model and serial numbers is available at the [CareFusion website](http://www.carefusion.com/customer-support/alerts-notices/?ref=foodconsumer.org). Refer to Figure 1 for the location of model and serial numbers.

CareFusion puts patient safety first and will continue to work diligently with customers and global regulatory authorities to resolve this issue in a timely manner. The company has notified customers of the recall with an urgent recall letter. The CareFusion Recall Support Center is contacting global customers via telephone to coordinate on-site AVEA ventilator correction in a timely, effective manner.

The FDA and other regulatory authorities have been notified.

**Instructions to Customers**

CareFusion does not require the return of affected AVEA ventilators.

In the interim, if an AVEA ventilator exhibits a sustained Extended High Ppeak or Circuit Occlusion alarm followed by the opening of the safety valve that cannot be cleared by powering the ventilator off and back on again, immediately remove the ventilator from service, provide alternate ventilation and contact CareFusion Technical Support at the contact information listed below to report the issue.

Customer inquiries related to this action should be addressed to the CareFusion Recall Support Center at 1.888.562.6018 or email [SupportCenter@carefusion.com](mailto:SupportCenter@carefusion.com) between the hours of 6:30 a.m. to 5:00 p.m. PDT.

For technical support or to report a problem, please contact CareFusion Technical Support at 1.800.231.2466 or email [Support.Vent.US@carefusion.com](mailto:Support.Vent.US@carefusion.com).

Adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax:

- Complete and submit the report Online: [www.fda.gov/medwatch/report.htm](http://www.fda.gov/medwatch/report.htm?ref=foodconsumer.org)
- Regular Mail or Fax: Download form [www.fda.gov/MedWatch/getforms.htm](http://www.fda.gov/MedWatch/getforms.htm?ref=foodconsumer.org) or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178.

Countries with AVEA ventilators affected by the global recall:

| **ISO Code** | **Country**              |
| ------------ | ------------------------ |
| AL           | Albania                  |
| AS           | American Samoa           |
| AR           | Argentina                |
| AU           | Australia                |
| AT           | Austria                  |
| BH           | Bahrain                  |
| BD           | Bangladesh               |
| BB           | Barbados                 |
| BE           | Belgium                  |
| BZ           | Belize                   |
| BM           | Bermuda                  |
| BO           | Bolivia                  |
| BA           | Bosnia and Heherzegowina |
| BR           | Brazil                   |
| CA           | Canada                   |
| CL           | Chile                    |
| CN           | China (People's Rep)     |
| CO           | Colombia                 |
| CR           | Costa Rica               |
| HR           | Croatia                  |
| CY           | Cyprus                   |
| CZ           | Czech (Rep)              |
| DO           | Dominican Republic       |
| EC           | Ecuador                  |
| EG           | Egypt                    |
| SV           | El Salvador              |
| SV           | El Salvador              |
| EE           | Estonia                  |
| FR           | France                   |
| GE           | Georgia                  |
| DE           | Germany                  |
| GB           | Great Britain            |
| GR           | Greece                   |
| GT           | Guatemala                |
| HN           | Honduras (Rep)           |
| HK           | Hong Kong, China         |
| HU           | Hungary (Rep)            |
| IN           | India                    |
| ID           | Indonesia                |
| IE           | Ireland                  |
| IL           | Israel                   |
| IT           | Italy                    |
| JP           | Japan                    |
| JO           | Jordan                   |
| KR           | Korea (Rep)              |
| KW           | Kuwait                   |
| LV           | Latvia                   |
| LB           | Lebanon                  |
| LY           | Libyan Jamahiriya        |
| LT           | Lithuania                |
| MY           | Malaysia                 |
| MX           | Mexico                   |
| MD           | Moldova (republic of)    |
| KR           | Korea (Rep)              |
| KW           | Kuwait                   |
| MA           | Morocco                  |
| NP           | Nepal                    |
| NL           | Netherlands              |
| Nl           | Nicaragua                |
| NO           | Norway                   |
| OM           | Oman                     |
| PK           | Pakistan                 |
| PS           | Palestine                |
| PA           | Panama (Rep)             |
| PY           | Paraguay                 |
| PE           | Peru                     |
| PH           | Philippines              |
| PL           | Poland                   |
| PT           | Portugal                 |
| PR           | Puerto Rico              |
| QA           | Qatar                    |
| RO           | Romania                  |
| RU           | Russian Federation       |
| RW           | Rwanda                   |
| SA           | Saudi Arabia             |
| SG           | Singapore                |
| SK           | Slovakia                 |
| SI           | Slovenia                 |
| ZA           | South Africa             |
| ES           | Spain                    |
| TW           | Taiwan                   |
| TH           | Thailand                 |
| TN           | Tunisia                  |
| TR           | Turkey                   |
| AE           | United Arab Emirates     |
| UK           | United Kingdom           |
| UY           | Uruguay                  |
| US           | United States            |
| VE           | Venezuela                |
| VN           | Vietnam                  |
| YE           | Yemen                    |

**About CareFusion**

CareFusion, a BD company, serves the health care industry with products and services that help hospitals measurably improve the safety and quality of care. The company develops industry-leading technologies including [Alaris®](http://www.carefusion.com/medical-products/infusion/?ref=foodconsumer.org) infusion pumps and [IV sets](http://www.carefusion.com/medical-products/infusion/iv-therapy/?ref=foodconsumer.org), [MaxPlus® and MaxZero™ IV connectors and sets](http://www.carefusion.com/medical-products/infusion/iv-sets-accessories/MaxPlus-clear.aspx?ref=foodconsumer.org), [Pyxis®](http://www.carefusion.com/medical-products/medication-management/?ref=foodconsumer.org) automated dispensing and [patient identification systems](http://www.carefusion.com/medical-products/medication-management/point-of-care-verification/?ref=foodconsumer.org), [AVEA](http://www.carefusion.com/medical-products/respiratory/ventilation/avea/?ref=foodconsumer.org)®, [LTV® series](http://www.carefusion.com/medical-products/respiratory/ventilation/ltv-series/?ref=foodconsumer.org) and [Air*Life*](http://www.carefusion.com/medical-products/respiratory/consumables/?ref=foodconsumer.org)® ventilation and respiratory products, [ChloraPrep](http://www.carefusion.com/medical-products/infection-prevention/skin-preparation/?ref=foodconsumer.org)® products, [MedMined](http://www.carefusion.com/medical-products/infection-prevention/surveillance-analytics/?ref=foodconsumer.org)® services for data mining surveillance, [V. Mueller](http://www.carefusion.com/medical-products/surgical/surgical-instruments/?ref=foodconsumer.org)® surgical instruments, and an extensive line of [products that support interventional medicine](http://www.carefusion.com/medical-products/interventional-procedures/?ref=foodconsumer.org). For more information please visit [www.carefusion.com](http://www.carefusion.com/?ls=PR&cmp=&WT.mc%5Fid=10,100,PR,CFN,CFN,Q3FY12Earnings,CFNboiler&ref=foodconsumer.org).

**About BD**

BD is a leading medical technology company that partners with customers and stakeholders to address many of the world’s most pressing and evolving health needs. Our innovative solutions are focused on improving medication management and patient safety; supporting infection prevention practices; equipping surgical and interventional procedures; improving drug delivery; aiding anesthesiology and respiratory care; advancing cellular research and applications; enhancing the diagnosis of infectious diseases and cancers; and supporting the management of diabetes. We are more than 45,000 associates in 50 countries who strive to fulfill our purpose of “*Helping all people live healthy lives*” by advancing the quality, accessibility, safety and affordability of healthcare around the world. In 2015, BD welcomed CareFusion and [its products](http://www.carefusion.com/medical-products/?ref=foodconsumer.org) into the BD family of solutions. For more information on BD, please visit [www.bd.com](file:///C:/Users/troy.kirkpatrick/Documents/News%20Releases/www.bd.com).

###

[RSS Feed for FDA Recalls Information](http://www.fda.gov/FDAgov/AboutFDA/ContactFDA/StayInformed/RSSFeeds/Recalls/rss.xml?ref=foodconsumer.org) \[[what’s this?](http://www.fda.gov/FDAgov/AboutFDA/ContactFDA/StayInformed/RSSFeeds/default.htm?ref=foodconsumer.org)\]

Photo: [Product Labels](http://www.fda.gov/Safety/Recalls/ucm448256.htm?ref=foodconsumer.org)